JGW-T1504450-v7

DocDB Home ]  [ Search ] [ Recent Changes ]

BS Actuator Design (including design for SRs and PRs)

Document #:
JGW-T1504450-v7
Document type:
T
Submitted by:
Yuta Michimura
Updated by:
Yuta Michimura
Document Created:
29 Oct 2015, 16:53
Contents Revised:
30 Nov 2015, 14:59
DB Info Revised:
16 Jun 2016, 15:28
Viewable by:
Modifiable by:
Other Versions:
JGW-T1504450-v6
04 Nov 2015, 16:13
JGW-T1504450-v5
04 Nov 2015, 14:13
JGW-T1504450-v4
04 Nov 2015, 14:02
JGW-T1504450-v3
02 Nov 2015, 16:51
JGW-T1504450-v2
29 Oct 2015, 17:14
JGW-T1504450-v1
29 Oct 2015, 16:53
Abstract:
Slides for the Special VIS+ISC meeting on BS magnets
http://gwwiki.icrr.u-tokyo.ac.jp/JGWwiki/KAGRA/subgroup/ifo/MIF/Minutes20151104

Includes magnet choice for SRs and PRs
Files in Document:
Notes and Changes:
== BS ==
* Design change summary
- Old design
-- BS TM: NdFeB 6mm dia, 3mm thick (0.0744 Am^2)
-- BS IM: NdFeB 10mm dia, 10mm thick (0.690 Am^2)

- New design
-- BS TM: SmCo 2mm dia, 3mm thick (http://gwwiki.icrr.u-tokyo.ac.jp/JGWwiki/KAGRA/subgroup/ifo/MIF/Minutes20151104)
-- BS IM: SmCo NO CHANGE in size (see [kagra-seis 01218], as opposed to http://gwwiki.icrr.u-tokyo.ac.jp/JGWwiki/KAGRA/Subgroups/VIS/VISMinutes20151113)

* Magnetic noise
- Old design (http://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/DocDB/ShowDocument?docid=4459)
-- Uses measured magnetic field fluctuation when few equipments
installed in the tunnel
-- Magnetic noise from TM: 1/100 of the displacement noise requirement (at 10Hz)
-- Magnetic noise from IM: 1/1000 of the displacement noise requirement (at 10Hz)

- New design
-- Use x50 of the magnetic field fluctuation measurement (~ use Virgo measurement)
-- Magnetic noise from TM: ~1/18 of the displacement noise requirement (at 10Hz)
-- Magnetic noise from IM: ~1/20 of the displacement noise requirement (at 10Hz)
-> OK!

* Saturation check
- Old design (http://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/DocDB/ShowDocument?docid=3453)
-- DAC output at lock acquisition for TM: 70 counts
-- DAC output during lock for IM: 3425 counts RMS

- New design
-- DAC output at lock acquisition for TM: ~630 counts
-- DAC output during lock for IM: ~3425 counts RMS
-> OK!


== SRs ==
* Design change summary
- Old design
-- SRs TM: NdFeB 6mm dia, 3mm thick (0.0744 Am^2)
-- SRs IM: NdFeB 10mm dia, 10mm thick (0.690 Am^2)

- New design
-- SRs TM: SmCo 2mm dia, 5mm thick (http://gwwiki.icrr.u-tokyo.ac.jp/JGWwiki/KAGRA/subgroup/ifo/MIF/Minutes20151104)
-- SRs IM: SmCo NO CHANGE in size

* Magnetic noise
- Old design (http://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/DocDB/ShowDocument?docid=4459)
-- Uses measured magnetic field fluctuation when few equipments
installed in the tunnel
-- Magnetic noise from TM: 1/30 of the displacement noise requirement (at 10Hz)
-- Magnetic noise from IM: 1/300 of the displacement noise requirement (at 10Hz)

- New design
-- Use x50 of the magnetic field fluctuation measurement (~ use Virgo measurement)
-- Magnetic noise from TM: ~1/3 of the displacement noise requirement (at 10Hz)
-- Magnetic noise from IM: ~1/6 of the displacement noise requirement (at 10Hz)
-> OK!

* Saturation check
- Old design (http://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/DocDB/ShowDocument?docid=3453)
-- DAC output at lock acquisition for TM: 1500 counts
-- DAC output during lock for IM: 2641 counts RMS

- New design
-- DAC output at lock acquisition for TM: ~8300 counts
-- DAC output during lock for IM: ~2641 counts RMS
-> OK!


== PRs ==
* Design change summary
- Old design
-- PRs TM: NdFeB 6mm dia, 3mm thick (0.0744 Am^2)
-- PRs IM: NdFeB 10mm dia, 10mm thick (0.690 Am^2)

- New design
-- PRs TM: SmCo NO CHANGE in size (http://gwwiki.icrr.u-tokyo.ac.jp/JGWwiki/KAGRA/subgroup/ifo/MIF/Minutes20151104)
-- PRs IM: SmCo NO CHANGE in size

* Magnetic noise
- Old design (http://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/DocDB/ShowDocument?docid=4459)
-- Uses measured magnetic field fluctuation when few equipments
installed in the tunnel
-- Magnetic noise from TM: 1/1e4 of the displacement noise requirement (at 10Hz)
-- Magnetic noise from IM: 1/1e5 of the displacement noise requirement (at 10Hz)

- New design
-- Use x50 of the magnetic field fluctuation measurement (~ use Virgo measurement)
-- Magnetic noise from TM: ~1/200 of the displacement noise requirement (at 10Hz)
-- Magnetic noise from IM: ~1/2000 of the displacement noise requirement (at 10Hz)
-> OK!

* Saturation check
- Old design (http://gwdoc.icrr.u-tokyo.ac.jp/cgi-bin/DocDB/ShowDocument?docid=3453)
-- DAC output at lock acquisition for TM: 1100000 counts
-- DAC output during lock for IM: 58883 counts RMS

- New design
-- DAC output at lock acquisition for TM: ~1100000 counts
-- DAC output during lock for IM: ~58883 counts RMS
-> CHANGE IN SUSPENSION OR COIL DRIVERS NEEDED!